Prolonged high-pressure is required for optimal stent deployment as assessed by optical coherence tomography.
- Source :
-
Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions [Catheter Cardiovasc Interv] 2014 Mar 01; Vol. 83 (4), pp. 521-7. Date of Electronic Publication: 2013 Feb 21. - Publication Year :
- 2014
-
Abstract
- Aims: Optimizing stent deployment is important for both acute- and long-term outcomes. High-pressure balloon inflation is the standard for coronary stent implantation. However, there is no standardized inflation protocol. We hypothesized that prolonged high-pressure balloon inflation until stabilization of inflation pressure is superior to a rapid inflation/deflation sequence for both stent expansion and strut apposition.<br />Methods and Results: A high-pressure rapid inflation/deflation sequence was deployed followed by angiography to assure no residual stenosis. Optical coherence tomography (OCT) was then performed followed by prolonged inflation until balloon pressure was stabilized for 30 sec using the same balloon at the same pressure as the rapid sequence. A second OCT run was then done. Thirteen thousand nine hundred thirteen stent struts were evaluated by OCT in 12 patients undergoing successful stenting. Stent expansion improved with prolonged (206 ± 115 sec) vs. rapid (28 ± 17 sec) inflation for both minimal stent diameter (3.0 ± 0.5 vs. 2.75 ± 0.44 mm, P < 0.0001) and area (7.83 ± 2.45 vs. 6.63 ± 1.85 mm(2) , P = 0.0003). The number of malapposed struts decreased (45 ± 41 vs. 88 ± 75, P = 0.005) as did the maximal malapposed strut distance (0.31 ± 0.2 vs. 0.43 ± 0.2 mm, P = 0.0001). Factors related to strut malapposition after rapid inflation included localized asymmetry in 67%, stent underexpansion in 75%, and stent undersizing in 67%.<br />Conclusions: These data demonstrate that prolonged inflation is superior to a rapid inflation/deflation technique for both stent expansion and strut apposition. We recommend for routine stent deployment a prolonged inflation protocol as described above to optimize stent deployment.<br /> (Copyright © 2012 Wiley Periodicals, Inc.)
- Subjects :
- Aged
Angioplasty, Balloon, Coronary adverse effects
Coronary Angiography
Coronary Stenosis diagnosis
Female
Humans
Male
Middle Aged
Predictive Value of Tests
Pressure
Prospective Studies
Prosthesis Design
Treatment Outcome
Angioplasty, Balloon, Coronary instrumentation
Angioplasty, Balloon, Coronary methods
Coronary Stenosis therapy
Drug-Eluting Stents
Tomography, Optical Coherence
Subjects
Details
- Language :
- English
- ISSN :
- 1522-726X
- Volume :
- 83
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions
- Publication Type :
- Academic Journal
- Accession number :
- 23436559
- Full Text :
- https://doi.org/10.1002/ccd.24724