Back to Search
Start Over
Performance of a high-sensitivity dedicated cardiac SPECT scanner for striatal uptake quantification in the brain based on analysis of projection data.
- Source :
-
Medical physics [Med Phys] 2013 Apr; Vol. 40 (4), pp. 042504. - Publication Year :
- 2013
-
Abstract
- Purpose: The authors have previously reported the advantages of high-sensitivity single-photon emission computed tomography (SPECT) systems for imaging structures located deep inside the brain. DaTscan (Isoflupane I-123) is a dopamine transporter (DaT) imaging agent that has shown potential for early detection of Parkinson disease (PD), as well as for monitoring progression of the disease. Realizing the full potential of DaTscan requires efficient estimation of striatal uptake from SPECT images. They have evaluated two SPECT systems, a conventional dual-head gamma camera with low-energy high-resolution collimators (conventional) and a dedicated high-sensitivity multidetector cardiac imaging system (dedicated) for imaging tasks related to PD.<br />Methods: Cramer-Rao bounds (CRB) on precision of estimates of striatal and background activity concentrations were calculated from high-count, separate acquisitions of the compartments (right striata, left striata, background) of a striatal phantom. CRB on striatal and background activity concentration were calculated from essentially noise-free projection datasets, synthesized by scaling and summing the compartment projection datasets, for a range of total detected counts. They also calculated variances of estimates of specific-to-nonspecific binding ratios (BR) and asymmetry indices from these values using propagation of error analysis, as well as the precision of measuring changes in BR on the order of the average annual decline in early PD.<br />Results: Under typical clinical conditions, the conventional camera detected 2 M counts while the dedicated camera detected 12 M counts. Assuming a normal BR of 5, the standard deviation of BR estimates was 0.042 and 0.021 for the conventional and dedicated system, respectively. For an 8% decrease to BR = 4.6, the signal-to-noise ratio were 6.8 (conventional) and 13.3 (dedicated); for a 5% decrease, they were 4.2 (conventional) and 8.3 (dedicated).<br />Conclusions: This implies that PD can be detected earlier with the dedicated system than with the conventional system; therefore, earlier identification of PD progression should be possible with the high-sensitivity dedicated SPECT camera.
- Subjects :
- Equipment Design
Equipment Failure Analysis
Heart anatomy & histology
Humans
Myocardium metabolism
Reproducibility of Results
Sensitivity and Specificity
Contrast Media pharmacokinetics
Corpus Striatum diagnostic imaging
Corpus Striatum metabolism
Image Enhancement instrumentation
Tomography, Emission-Computed, Single-Photon instrumentation
Subjects
Details
- Language :
- English
- ISSN :
- 2473-4209
- Volume :
- 40
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- Medical physics
- Publication Type :
- Academic Journal
- Accession number :
- 23556920
- Full Text :
- https://doi.org/10.1118/1.4794488