Back to Search Start Over

Determination of erythromycin base and 2'-acetylerythromycin in human plasma using high-performance liquid chromatography with electrochemical detection.

Authors :
Laakso S
Scheinin M
Anttila M
Source :
Journal of chromatography [J Chromatogr] 1990 Apr 06; Vol. 526 (2), pp. 475-86.
Publication Year :
1990

Abstract

A high-performance liquid chromatographic method for the determination of plasma concentrations of erythromycin base and 2'-acetylerythromycin, an ester prodrug of erythromycin, is described. tert.-Butyl methyl ether extracts of 1-ml plasma samples (pH 10) were chromatographed on a C18 reversed-phase column. A three-electrode coulometric detector (oxidation potentials +0.65 and +0.85 V) was used for quantitation. Oleandomycin was used as an internal standard. The method has good precision and accuracy, is linear in the range 0.25-7.5 mg/l and has proved to be suitable for pharmacokinetic studies in humans. Correlation with a microbiological assay was satisfactory (r greater than or equal to 0.95), but the chromatographic method gave ca. 30% higher values.

Details

Language :
English
Volume :
526
Issue :
2
Database :
MEDLINE
Journal :
Journal of chromatography
Publication Type :
Academic Journal
Accession number :
2361986
Full Text :
https://doi.org/10.1016/s0378-4347(00)82529-3