Intraobserver reliability of contact pachymetry in children.

Authors :
Weise KK
Kaminski B
Melia M
Repka MX
Bradfield YS
Davitt BV
Johnson DA
Kraker RT
Manny RE
Matta NS
Schloff S
Source :
Journal of AAPOS : the official publication of the American Association for Pediatric Ophthalmology and Strabismus [J AAPOS] 2013 Apr; Vol. 17 (2), pp. 144-8.
Publication Year :
2013

Abstract

Background: Central corneal thickness (CCT) is an important measurement in the treatment and management of pediatric glaucoma and potentially of refractive error, but data regarding reliability of CCT measurement in children are limited. The purpose of this study was to evaluate the reliability of CCT measurement with the use of handheld contact pachymetry in children.<br />Methods: We conducted a multicenter intraobserver test-retest reliability study of more than 3,400 healthy eyes in children aged from newborn to 17 years by using a handheld contact pachymeter (Pachmate DGH55; DGH Technology Inc, Exton, PA) in 2 clinical settings--with the use of topical anesthesia in the office and with the patient under general anesthesia in a surgical facility.<br />Results: The overall standard error of measurement, including only measurements with standard deviation ≤5 μm, was 8 μm; the corresponding coefficient of repeatability, or limits within which 95% of test-retest differences fell, was ±22.3 μm. However, standard error of measurement increased as CCT increased, from 6.8 μm for CCT less than 525 μm, to 12.9 μm for CCT 625 μm and greater. The standard error of measurement including measurements with standard deviation >5 μm was 10.5 μm. Age, sex, race/ethnicity group, and examination setting did not influence the magnitude of test-retest differences.<br />Conclusions: CCT measurement reliability in children via the Pachmate DGH55 handheld contact pachymeter is similar to that reported for adults. Because thicker CCT measurements are less reliable than thinner measurements, a second measure may be helpful when the first exceeds 575 μm. Reliability is also improved by disregarding measurements with instrument-reported standard deviations >5 μm.<br /> (Copyright © 2013 American Association for Pediatric Ophthalmology and Strabismus. Published by Mosby, Inc. All rights reserved.)

Details

Language :
English
ISSN :
1528-3933
Volume :
17
Issue :
2
Database :
MEDLINE
Journal :
Journal of AAPOS : the official publication of the American Association for Pediatric Ophthalmology and Strabismus
Publication Type :
Academic Journal
Accession number :
23622447
Full Text :
https://doi.org/10.1016/j.jaapos.2012.11.005