Improved porous silicon microarray based prostate specific antigen immunoassay by optimized surface density of the capture antibody.

Authors :
Lee S
Kim S
Malm J
Jeong OC
Lilja H
Laurell T
Source :
Analytica chimica acta [Anal Chim Acta] 2013 Sep 24; Vol. 796, pp. 108-14. Date of Electronic Publication: 2013 Jul 03.
Publication Year :
2013

Abstract

Enriching the surface density of immobilized capture antibodies enhances the detection signal of antibody sandwich microarrays. In this study, we improved the detection sensitivity of our previously developed P-Si (porous silicon) antibody microarray by optimizing concentrations of the capturing antibody. We investigated immunoassays using a P-Si microarray at three different capture antibody (PSA - prostate specific antigen) concentrations, analyzing the influence of the antibody density on the assay detection sensitivity. The LOD (limit of detection) for PSA was 2.5 ng mL(-1), 80 pg mL(-1), and 800 fg mL(-1) when arraying the PSA antibody, H117 at the concentration 15 μg mL(-1), 35 μg mL(-1), and 154 μg mL(-1), respectively. We further investigated PSA spiked into human female serum in the range of 800 fg mL(-1) to 500 ng mL(-1). The microarray showed a LOD of 800 fg mL(-1) and a dynamic range of 800 fg mL(-1) to 80 ng mL(-1) in serum spiked samples.<br /> (Copyright © 2013 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-4324
Volume :
796
Database :
MEDLINE
Journal :
Analytica chimica acta
Publication Type :
Academic Journal
Accession number :
24016590
Full Text :
https://doi.org/10.1016/j.aca.2013.06.041