Improved porous silicon microarray based prostate specific antigen immunoassay by optimized surface density of the capture antibody.
- Source :
-
Analytica chimica acta [Anal Chim Acta] 2013 Sep 24; Vol. 796, pp. 108-14. Date of Electronic Publication: 2013 Jul 03. - Publication Year :
- 2013
-
Abstract
- Enriching the surface density of immobilized capture antibodies enhances the detection signal of antibody sandwich microarrays. In this study, we improved the detection sensitivity of our previously developed P-Si (porous silicon) antibody microarray by optimizing concentrations of the capturing antibody. We investigated immunoassays using a P-Si microarray at three different capture antibody (PSA - prostate specific antigen) concentrations, analyzing the influence of the antibody density on the assay detection sensitivity. The LOD (limit of detection) for PSA was 2.5 ng mL(-1), 80 pg mL(-1), and 800 fg mL(-1) when arraying the PSA antibody, H117 at the concentration 15 μg mL(-1), 35 μg mL(-1), and 154 μg mL(-1), respectively. We further investigated PSA spiked into human female serum in the range of 800 fg mL(-1) to 500 ng mL(-1). The microarray showed a LOD of 800 fg mL(-1) and a dynamic range of 800 fg mL(-1) to 80 ng mL(-1) in serum spiked samples.<br /> (Copyright © 2013 Elsevier B.V. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1873-4324
- Volume :
- 796
- Database :
- MEDLINE
- Journal :
- Analytica chimica acta
- Publication Type :
- Academic Journal
- Accession number :
- 24016590
- Full Text :
- https://doi.org/10.1016/j.aca.2013.06.041