Antibody-free detection of human chorionic gonadotropin by use of liquid crystals.

Authors :
Ding X
Yang KL
Source :
Analytical chemistry [Anal Chem] 2013 Nov 19; Vol. 85 (22), pp. 10710-6. Date of Electronic Publication: 2013 Nov 06.
Publication Year :
2013

Abstract

Human chorionic gonadotropin (hCG) is an important biomarker for the diagnosis of pregnancy and cancers. In this study, we report an antibody-free and label-free mechanism for detecting hCG. To replace enzyme-labeled antibodies, we use a short oligopeptide as an hCG receptor to bind hCG. The short oligopeptide sequence, (N-)PPLRINRHILTR(-C), is identified after 5 rounds of screening by use of a phage library. After binding, liquid crystal (LC) is used to transduce the binding event into optical signals. The captured hCG can disrupt a thin layer (~6 μm) of LC covered on the surface. Depending on the initial concentration of hCG, LC gives distinct optical signals visible to the naked eye. The limit of detection (LOD) for this method is approximately 1 IU/mL (2 nM) in both phosphate-buffered saline and urine samples, and only 0.6 μL of hCG solution is required. This means that as little as 45.5 pg of hCG can be detected by this method. Compared to other detection methods for hCG, this detection method does not require the use of antibody and is label-free. It has the potential to become a portable diagnostic kit for hCG.

Details

Language :
English
ISSN :
1520-6882
Volume :
85
Issue :
22
Database :
MEDLINE
Journal :
Analytical chemistry
Publication Type :
Academic Journal
Accession number :
24147645
Full Text :
https://doi.org/10.1021/ac400732n