- A novel chemiluminescence (CL) system with diperiodatonickelate (IV) (DPN) was developed for the determination of adrenaline for the first time. The possible CL emission mechanism was briefly discussed by comparing the fluorescence emission spectra with CL spectra. Under the optimum conditions, the relative CL intensity was linear over the concentration of AD ranging from 1.0×10(-7) to 1.0×10(-5) g mL(-1) with a detection limit of 4.0×10(-8) g mL(-1) (3σ). And the relative standard deviation was 3.7% for 2.0×10(-6) g mL(-1) AD (n=11). The developed method has been successfully applied to the determination of AD in pharmaceutical preparations.<br /> (Copyright © 2013 Elsevier B.V. All rights reserved.)