Detection of carbapenemases by real-time PCR and melt curve analysis on the BD Max system.

Authors :
Hofko M
Mischnik A
Kaase M
Zimmermann S
Dalpke AH
Source :
Journal of clinical microbiology [J Clin Microbiol] 2014 May; Vol. 52 (5), pp. 1701-4. Date of Electronic Publication: 2014 Feb 12.
Publication Year :
2014

Abstract

We developed a multiplex SYBR green real-time PCR for the BD Max instrument (BD Diagnostics, Sparks, MD) to detect a panel of carbapenemases. The assay was evaluated with 152 consecutive isolates sent to the German National Reference Laboratory, and 65/65 of the carbapenemase-positive and 87/87 of the carbapenemase-negative strains were identified correctly.