Detection of carbapenemases by real-time PCR and melt curve analysis on the BD Max system.
- Source :
-
Journal of clinical microbiology [J Clin Microbiol] 2014 May; Vol. 52 (5), pp. 1701-4. Date of Electronic Publication: 2014 Feb 12. - Publication Year :
- 2014
-
Abstract
- We developed a multiplex SYBR green real-time PCR for the BD Max instrument (BD Diagnostics, Sparks, MD) to detect a panel of carbapenemases. The assay was evaluated with 152 consecutive isolates sent to the German National Reference Laboratory, and 65/65 of the carbapenemase-positive and 87/87 of the carbapenemase-negative strains were identified correctly.
Details
- Language :
- English
- ISSN :
- 1098-660X
- Volume :
- 52
- Issue :
- 5
- Database :
- MEDLINE
- Journal :
- Journal of clinical microbiology
- Publication Type :
- Academic Journal
- Accession number :
- 24523474
- Full Text :
- https://doi.org/10.1128/JCM.00373-14