Back to Search Start Over

Feasibility of non-invasive temperature estimation by the assessment of the average gray-level content of B-mode images.

Authors :
Teixeira CA
Alvarenga AV
Cortela G
von Krüger MA
Pereira WC
Source :
Ultrasonics [Ultrasonics] 2014 Aug; Vol. 54 (6), pp. 1692-702. Date of Electronic Publication: 2014 Mar 01.
Publication Year :
2014

Abstract

This paper assesses the potential of the average gray-level (AVGL) from ultrasonographic (B-mode) images to estimate temperature changes in time and space in a non-invasive way. Experiments were conducted involving a homogeneous bovine muscle sample, and temperature variations were induced by an automatic temperature regulated water bath, and by therapeutic ultrasound. B-mode images and temperatures were recorded simultaneously. After data collection, regions of interest (ROIs) were defined, and the average gray-level variation computed. For the selected ROIs, the AVGL-Temperature relation were determined and studied. Based on uniformly distributed image partitions, two-dimensional temperature maps were developed for homogeneous regions. The color-coded temperature estimates were first obtained from an AVGL-Temperature relation extracted from a specific partition (where temperature was independently measured by a thermocouple), and then extended to the other partitions. This procedure aimed to analyze the AVGL sensitivity to changes not only in time but also in space. Linear and quadratic relations were obtained depending on the heating modality. We found that the AVGL-Temperature relation is reproducible over successive heating and cooling cycles. One important result was that the AVGL-Temperature relations extracted from one region might be used to estimate temperature in other regions (errors inferior to 0.5 °C) when therapeutic ultrasound was applied as a heating source. Based on this result, two-dimensional temperature maps were developed when the samples were heated in the water bath and also by therapeutic ultrasound. The maps were obtained based on a linear relation for the water bath heating, and based on a quadratic model for the therapeutic ultrasound heating. The maps for the water bath experiment reproduce an acceptable heating/cooling pattern, and for the therapeutic ultrasound heating experiment, the maps seem to reproduce temperature profiles consistent with the pressure field of the transducer, and in agreement with temperature maps developed by COMSOL®MultiPhysics simulations.<br /> (Copyright © 2014 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1874-9968
Volume :
54
Issue :
6
Database :
MEDLINE
Journal :
Ultrasonics
Publication Type :
Academic Journal
Accession number :
24630851
Full Text :
https://doi.org/10.1016/j.ultras.2014.02.021