Back to Search
Start Over
Evaluation of the Bruker Biotyper matrix-assisted laser desorption ionization-time of flight mass spectrometry system for identification of blood isolates of Acinetobacter species.
- Source :
-
Journal of clinical microbiology [J Clin Microbiol] 2014 Aug; Vol. 52 (8), pp. 3095-100. Date of Electronic Publication: 2014 Jun 04. - Publication Year :
- 2014
-
Abstract
- Matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) (Bruker Biotyper) was able to accurately identify 98.6% (142/144) of Acinetobacter baumannii isolates, 72.4% (63/87) of A. nosocomialis isolates, and 97.6% (41/42) of A. pittii isolates. All Acinetobacter junii, A. ursingii, A. johnsonii, and A. radioresistens isolates (n = 28) could also be identified correctly by Bruker Biotyper.<br /> (Copyright © 2014, American Society for Microbiology. All Rights Reserved.)
- Subjects :
- Acinetobacter chemistry
Acinetobacter isolation & purification
Humans
Acinetobacter classification
Acinetobacter Infections diagnosis
Acinetobacter Infections microbiology
Bacteremia diagnosis
Bacteremia microbiology
Bacteriological Techniques methods
Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization methods
Subjects
Details
- Language :
- English
- ISSN :
- 1098-660X
- Volume :
- 52
- Issue :
- 8
- Database :
- MEDLINE
- Journal :
- Journal of clinical microbiology
- Publication Type :
- Academic Journal
- Accession number :
- 24899038
- Full Text :
- https://doi.org/10.1128/JCM.01233-14