Studies on quantitative phosphopeptide analysis by matrix-assisted laser desorption/ionization mass spectrometry without label, chromatography or calibration curves.

Authors :
Ho HP
Rathod P
Louis M
Tada CK
Rahaman S
Mark KJ
Leng J
Dana D
Kumar S
Lichterfeld M
Chang EJ
Source :
Rapid communications in mass spectrometry : RCM [Rapid Commun Mass Spectrom] 2014 Dec 30; Vol. 28 (24), pp. 2681-9.
Publication Year :
2014

Abstract

Rationale: Matrix-assisted laser desorption/ionization (MALDI) mass spectrometry combined with isotope labeling methods are effective for protein and peptide quantification, but limited in their multiplexing capacity, cost-effectiveness and dynamic range. This study investigates MALDI-MS-based quantification of peptide phosphorylation without labeling, and aims to overcome the shot-to-shot variability of MALDI using a mathematical transformation and extended data acquisition times.<br />Methods: A linear relationship between the reciprocal of phosphopeptide mole fraction and the reciprocal of phosphorylated-to-unphosphorylated signal ratio is derived, and evaluated experimentally using three separate phosphopeptide systems containing phosphorylated serine, threonine and tyrosine residues: mixtures of phosphopeptide and its des-phospho-analog with known stoichiometry measured by vacuum MALDI-linear ion trap mass spectrometry and fit to the linear model. The model is validated for quantifying in vitro phosphorylation assays with inhibition studies on Cdk2/cyclinA.<br />Results: Dynamic range of picomoles to femtomoles, good accuracy (deviations of 1.5-3.0% from expected values) and reproducibility (relative standard deviation (RSD) = 4.3-6.3%) are achieved. Inhibition of cyclin-dependent kinase phosphorylation by the classical inhibitors olomoucine and r-roscovitine was evaluated and IC50 values found to be in agreement with reported literature values. These results, achieved with single-point calibration, without isotope or chromatography, compare favorably to those arrived at using isotope dilution (p > 0.5 for accuracy).<br />Conclusions: The mathematical relationship derived here can be applied to a method that we term Double Reciprocal Isotope-free Phosphopeptide Quantification (DRIP-Q), as a strategy for quantification of in vitro phosphorylation assays, the first MALDI-based, isotope- and calibration curve-free method of its type. These results also pave the way for further systematic studies investigating the effect of peptide composition and experimental conditions on quantitative, label-free MALDI.<br /> (Copyright © 2014 John Wiley & Sons, Ltd.)

Details

Language :
English
ISSN :
1097-0231
Volume :
28
Issue :
24
Database :
MEDLINE
Journal :
Rapid communications in mass spectrometry : RCM
Publication Type :
Academic Journal
Accession number :
25380489
Full Text :
https://doi.org/10.1002/rcm.7063