Back to Search Start Over

X-ray crystal spectrometer upgrade for ITER-like wall experiments at JET.

Authors :
Shumack AE
Rzadkiewicz J
Chernyshova M
Jakubowska K
Scholz M
Byszuk A
Cieszewski R
Czarski T
Dominik W
Karpinski L
Kasprowicz G
Pozniak K
Wojenski A
Zabolotny W
Conway NJ
Dalley S
Figueiredo J
Nakano T
Tyrrell S
Zastrow KD
Zoita V
Source :
The Review of scientific instruments [Rev Sci Instrum] 2014 Nov; Vol. 85 (11), pp. 11E425.
Publication Year :
2014

Abstract

The high resolution X-Ray crystal spectrometer at the JET tokamak has been upgraded with the main goal of measuring the tungsten impurity concentration. This is important for understanding impurity accumulation in the plasma after installation of the JET ITER-like wall (main chamber: Be, divertor: W). This contribution provides details of the upgraded spectrometer with a focus on the aspects important for spectral analysis and plasma parameter calculation. In particular, we describe the determination of the spectrometer sensitivity: important for impurity concentration determination.

Details

Language :
English
ISSN :
1089-7623
Volume :
85
Issue :
11
Database :
MEDLINE
Journal :
The Review of scientific instruments
Publication Type :
Academic Journal
Accession number :
25430332
Full Text :
https://doi.org/10.1063/1.4891182