Conceptual design of a divertor Thomson scattering diagnostic for NSTX-U.

Authors :
McLean AG
Soukhanovskii VA
Allen SL
Carlstrom TN
LeBlanc BP
Ono M
Stratton BC
Source :
The Review of scientific instruments [Rev Sci Instrum] 2014 Nov; Vol. 85 (11), pp. 11E825.
Publication Year :
2014

Abstract

A conceptual design for a divertor Thomson scattering (DTS) diagnostic has been developed for the NSTX-U device to operate in parallel with the existing multipoint Thomson scattering system. Higher projected peak heat flux in NSTX-U will necessitate application of advanced magnetics geometries and divertor detachment. Interpretation and modeling of these divertor scenarios will depend heavily on local measurement of electron temperature, Te, and density, ne, which DTS provides in a passive manner. The DTS design for NSTX-U adopts major elements from the successful DIII-D DTS system including 7-channel polychromators measuring Te to 0.5 eV. If implemented on NSTX-U, the divertor TS system would provide an invaluable diagnostic for the boundary program to characterize the edge plasma.

Details

Language :
English
ISSN :
1089-7623
Volume :
85
Issue :
11
Database :
MEDLINE
Journal :
The Review of scientific instruments
Publication Type :
Academic Journal
Accession number :
25430390
Full Text :
https://doi.org/10.1063/1.4894001