Bioabsorbable interference screw failure in anterior cruciate ligament reconstruction: A case series and review of the literature.

Authors :
Watson JN
McQueen P
Kim W
Hutchinson MR
Source :
The Knee [Knee] 2015 Jun; Vol. 22 (3), pp. 256-61. Date of Electronic Publication: 2015 Mar 17.
Publication Year :
2015

Abstract

Background: To report a case series of failures of bioabsorbable interference screws with possible identification of a novel failure mechanism.<br />Methods: A retrospective review of ACL reconstructions by the senior author utilizing BioCompositeā„¢ Interference Screws (Arthrex, Inc., Naples, FL) was performed. Complications related to screw placement, including fracture, breakage or bending were examined. Our rate and methods of failure were compared to those quoted in the current literature.<br />Results: Eighty-seven patients of average age 23.8 years met inclusion criteria. There were eight screw failures in six patients, with femoral failure in seven and tibial failure in one. The femoral screw fractured halfway between the tip and head in five, while the head of the screw broke in one and the screw bent in another. In the case of tibial interference screw fracture, failure occurred halfway between the tip and head. The insertion device that was used was replaced after recognition of material deformation and considered a potential contributor to the breakage risk as no further screw failures have occurred since.<br />Conclusions: We demonstrate a unique failure mechanism of bio-absorbable interference screws. In each case, the reconstruction was salvaged. Regular inspection of materials and implants can ensure optimal outcomes and decrease complications intra-operatively.<br /> (Copyright © 2015 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-5800
Volume :
22
Issue :
3
Database :
MEDLINE
Journal :
The Knee
Publication Type :
Academic Journal
Accession number :
25795545
Full Text :
https://doi.org/10.1016/j.knee.2015.02.015