Removing Ambiguity Caused by T2 Shine-through using Weighted Diffusion Subtraction (WDS).
- Source :
-
Magnetic resonance in medical sciences : MRMS : an official journal of Japan Society of Magnetic Resonance in Medicine [Magn Reson Med Sci] 2016; Vol. 15 (1), pp. 146-8. Date of Electronic Publication: 2015 Jun 23. - Publication Year :
- 2016
-
Abstract
- We propose a novel image processing technique that combines images routinely acquired with low and high b values to create a single image that contains clinically useful information without the ambiguity of T2 shine-through. The contrast of resulting images is similar to that of a T2 image, but the signals of pixels with low apparent diffusion coefficient (ADC) values are inverted. The proposed technique takes the threshold ADC value as the one adjustable parameter.