Type IIIb Endoleak With the Endurant Stent-Graft.
- Source :
-
Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists [J Endovasc Ther] 2016 Feb; Vol. 23 (1), pp. 229-32. Date of Electronic Publication: 2015 Nov 12. - Publication Year :
- 2016
-
Abstract
- Purpose: To report 2 cases of type IIIb endoleak with the Endurant stent-graft and postulate the cause for the events.<br />Case Report: A type IIIb endoleak was diagnosed at open conversion for a ruptured aneurysm 4 years after implantation of an Endurant stent-graft. In the other case, the endoleak was diagnosed at angiography 4 years after the Endurant stent-graft was implanted; the stent-graft was relined. In both cases the fabric hole was in the body of the stent-graft at the level of the top of the contralateral limb.<br />Conclusion: The cause of the type IIIb endoleaks in these cases was fabric erosion likely due to interaction between the bare metal at the top of the contralateral limb and the fabric of the stent-graft body.<br /> (© The Author(s) 2015.)
- Subjects :
- Aged
Aged, 80 and over
Aortic Aneurysm, Abdominal diagnosis
Aortography methods
Blood Vessel Prosthesis Implantation adverse effects
Endoleak diagnosis
Endoleak surgery
Endovascular Procedures adverse effects
Humans
Male
Prosthesis Design
Reoperation
Tomography, X-Ray Computed
Treatment Outcome
Aortic Aneurysm, Abdominal surgery
Blood Vessel Prosthesis
Blood Vessel Prosthesis Implantation instrumentation
Endoleak etiology
Endovascular Procedures instrumentation
Stents
Subjects
Details
- Language :
- English
- ISSN :
- 1545-1550
- Volume :
- 23
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists
- Publication Type :
- Academic Journal
- Accession number :
- 26564914
- Full Text :
- https://doi.org/10.1177/1526602815617477