Back to Search Start Over

[Determination of total Bromine in urine by inductively coupled plasma mass spectrometry (ICP-MS)].

Authors :
Zhou C
Zhu H
Liu D
Miao R
Ying L
Zhu B
Source :
Zhonghua lao dong wei sheng zhi ye bing za zhi = Zhonghua laodong weisheng zhiyebing zazhi = Chinese journal of industrial hygiene and occupational diseases [Zhonghua Lao Dong Wei Sheng Zhi Ye Bing Za Zhi] 2015 May; Vol. 33 (5), pp. 396-7.
Publication Year :
2015

Abstract

Objective: To establish a method to determine total bromine in urine.<br />Method: Diluted urine samples were directly introduced into ICP-MS then quantized by standard curve.<br />Result: Total bromine in urine was linear within 1.0~50 mg/L with r > 0.999, When spiked at a concentration of 0.020 mg/L, 0.050 mg/L, 0.150 mg/L, the recovery was 95%~98%, intra-assay precision was 1.4% 3.2%, inter-assay precision was 3.4% to 5.0%. Urine could store in -20 °C refrigerator 3 months without any bromine loss.<br />Conclusion: Using ICP-MS to determine the urinary total bromine, the method is fast, accurate, wide linear range of features, could meet with the requirement of Part 5 of occupational health standards guide: Method determination of chemical substances in biological materials (GBZ/T 210.5-2008), a strong competitive advantage in a wide range of survey, suitable for promotion.

Details

Language :
Chinese
ISSN :
1001-9391
Volume :
33
Issue :
5
Database :
MEDLINE
Journal :
Zhonghua lao dong wei sheng zhi ye bing za zhi = Zhonghua laodong weisheng zhiyebing zazhi = Chinese journal of industrial hygiene and occupational diseases
Publication Type :
Academic Journal
Accession number :
26653245