Abstract
- The paper describes the measurements performed at IFIN-HH regarding the creation of a Romanian (124)I standard, consisting of: absolute standardization of the solution by the application of the 4πβ(PC)-γ coincidence method; Calibration of the CENTRONIC IG12/20A ionization chamber with a standardised solution and comparison with a calculated efficiency; γ-ray spectrometry activity measurement and determination of the impurity levels; Comparison of the results of the three methods.<br /> (Copyright © 2015 Elsevier Ltd. All rights reserved.)