Comparative analysis of iterative reconstruction algorithms with resolution recovery and time of flight modeling for 18 F-FDG cardiac PET: A multi-center phantom study.

Authors :
Matheoud R
Lecchi M
Lizio D
Scabbio C
Marcassa C
Leva L
Del Sole A
Rodella C
Indovina L
Bracco C
Brambilla M
Zoccarato O
Source :
Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology [J Nucl Cardiol] 2017 Jun; Vol. 24 (3), pp. 1036-1045. Date of Electronic Publication: 2016 Jan 12.
Publication Year :
2017

Abstract

Background: The purpose of this study was to evaluate the image quality in cardiac <superscript>18</superscript> F-FDG PET using the time of flight (TOF) and/or point spread function (PSF) modeling in the iterative reconstruction (IR).<br />Methods: Three scanners and an anthropomorphic cardiac phantom with an insert simulating a transmural defect (TD) were used. Two sets of scans (with/without TD) were acquired, and four reconstruction schemes were considered: (1) IR; (2) IR + PSF, (3) IR + TOF, and (4) IR + TOF + PSF. LV wall thickness (FWHM), contrast between LV wall and inner chamber (C <subscript>IC</subscript> ), and TD contrast in LV wall (C <subscript>TD</subscript> ) were evaluated.<br />Results: Tests of the reconstruction protocols showed a decrease in FWHM from IR (13 mm) to IR + PSF (11 mm); an increase in the C <subscript>IC</subscript> from IR (65%) to IR + PSF (71%) and from IR + TOF (72%) to IR + TOF + PSF (77%); and an increase in the C <subscript>TD</subscript> from IR + PSF (72%) to IR + TOF (75%) and to IR + TOF + PSF (77%). Tests of the scanner/software combinations showed a decrease in FWHM from Gemini_TF (13 mm) to Biograph_mCT (12 mm) and to Discovery_690 (11 mm); an increase in the C <subscript>IC</subscript> from Gemini_TF (65%) to Biograph_mCT (73%) and to Discovery_690 (75%); and an increase in the C <subscript>TD</subscript> from Gemini_TF/Biograph_mCT (72%) to Discovery_690 (77%).<br />Conclusion: The introduction of TOF and PSF increases image quality in cardiac <superscript>18</superscript> F-FDG PET. The scanner/software combinations exhibit different performances, which should be taken into consideration when making cross comparisons.

Details

Language :
English
ISSN :
1532-6551
Volume :
24
Issue :
3
Database :
MEDLINE
Journal :
Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology
Publication Type :
Academic Journal
Accession number :
26758376
Full Text :
https://doi.org/10.1007/s12350-015-0385-z