One-year clinical outcome after ablation with a novel multipolar irrigated ablation catheter for treatment of atrial fibrillation: potential implications for clinical use.

Authors :
Wakili R
Siebermair J
Fichtner S
Sinner MF
Klocker E
Olesch L
Hilberath JN
Sarai S
Clauss S
Sattler S
Kääb S
Estner HL
Source :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology [Europace] 2016 Aug; Vol. 18 (8), pp. 1170-8. Date of Electronic Publication: 2016 Jan 11.
Publication Year :
2016

Abstract

Aims: Pulmonary vein isolation (PVI) is an established therapy for atrial fibrillation (AF). However, PVI remains a time-consuming procedure. A novel multipolar irrigated radiofrequency (RF) ablation catheter (nMARQ™) is aiming to improve PVI. We investigated the influence on procedural parameters and assessed clinical outcomes after PVI using this novel catheter.<br />Methods and Results: Fifty-eight consecutive patients with paroxysmal AF were equally allocated (n = 29/group) to PVI treatment with (i) the novel multipolar ablation catheter (nMARQ™) and (ii) a standard single-tip ablation catheter (SAC). Study endpoints included procedure time, fluoroscopy time, radiation dose, RF time, number of energy applications, and clinical outcome defined as freedom from AF after a single procedure. Successful PVI was confirmed by a separate circular, multipolar mapping catheter in all patients treated with the nMARQ™. Pulmonary vein isolation was achieved in 100% in the SAC group. In the nMARQ™ group, PVI was suggested in all patients. However, confirmatory mapping revealed persistent pulmonary vein (PV) conduction in 19 out of 29 nMARQ™ patients. These patients underwent further ablation, which still failed to achieve PVI in 5 of the 29 nMARQ™ patients, mainly due to significant temperature rise in the oesophagus and device-related limitations reaching the right inferior PV. Mean fluoroscopy time (31 ± 12 vs. 23 ± 10 min, P < 0.05) and (132 ± 37 vs. 109 ± 30 min, P < 0.05) were longer in nMARQ™ vs. SAC patients. Radiofrequency time was shorter in nMARQ™ vs. SAC group (21 ± 9 vs. 35 ± 12 min, P < 0.001). Radiation dose and the number of energy applications did not differ between both groups. Clinical outcome analysis revealed no significant differences (nMARQ™: 72 vs. SAC: 72%) after a mean follow-up of 373 ± 278 days.<br />Conclusion: The use of the nMARQ™ catheter is associated with important device-related limitations to achieve successful PVI. However, clinical outcomes were equivalent in nMARQ™- and SAC-treated patients.<br /> (Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2016. For permissions please email: journals.permissions@oup.com.)

Details

Language :
English
ISSN :
1532-2092
Volume :
18
Issue :
8
Database :
MEDLINE
Journal :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
Publication Type :
Academic Journal
Accession number :
26759125
Full Text :
https://doi.org/10.1093/europace/euv349