Back to Search Start Over

Impact of Magnetic Field on Pressures of Programmable Cerebrospinal Fluid Shunts: An Experimental Study.

Authors :
Altun I
Yuksel KZ
Mert T
Source :
Turkish neurosurgery [Turk Neurosurg] 2017; Vol. 27 (6), pp. 935-941.
Publication Year :
2017

Abstract

Aim: To investigate whether programmable cerebrospinal fluid (CSF) shunts are influenced by exposure to the magnetic field and to compare the effects of magnetic field in 4 different brands of programmable CSF shunts.<br />Material and Methods: This experimental study was performed in the laboratory using a novel design of magnetic field. Four types of programmable CSF shunts (Miethke®, Medtronic®, Sophysa® and Codman®Hakim®) were exposed to the magnetic field generated by an apparatus consisting of Helmholtz coil for 5 minutes. In every CSF shunt, initial pressures were adjusted to 110 mm H2O and pressures after exposure to magnetic field were noted. These measurements were implemented at frequencies of 5 Hz, 20 Hz, 30 Hz, 40 Hz, 60 Hz and 80 Hz. In each type, three shunts were utilized and evaluations were made twice for every shunt.<br />Results: At 5, 30, 40 and 60 Hz, Groups 1, 2 and 3 had significantly higher average pressures than Group 4. At 20 and 80 Hz, Groups 1 and 2 had notably different pressure values than Groups 3 and 4. Group 3 displayed the highest pressure, while Group 4 demonstrated the lowest pressure.<br />Conclusion: Exposure to magnetic fields may affect the pressures of programmable CSF shunts. However, further controlled, clinical trials are warranted to elucidate the in-vivo effects of magnetic field exposure.

Details

Language :
English
ISSN :
2651-5032
Volume :
27
Issue :
6
Database :
MEDLINE
Journal :
Turkish neurosurgery
Publication Type :
Academic Journal
Accession number :
27476923
Full Text :
https://doi.org/10.5137/1019-5149.JTN.16795-15.2