Abstract

Molecular beacons carrying JOE dye (4',5'-dichloro-2',7'-dimethoxy-6-carboxyfluorescein) on a rigid or flexible linker and one or two BHQ1 quenchers have been prepared and tested in real-time PCR using Fusarium avenaceum elongation factor 1α DNA template. The probes were different in their structures (loop size and stem length), linkers for dye attachment (6-aminohexanol or trans-4-aminocyclohexanol), quencher composition (single and double BHQ1) to elucidate the influence of all these features. Fluorogenic properties of the probes were studied and compared to those of FAM (fluorescein)-based probes. All the factors - stem length, JOE vs FAM, rigid vs flexible linker, single vs double quencher - appeared to play a considerable role in the probe's fluorescent properties and determine the usability of the probe at two different temperatures of fluorescence detection (55°С and 64°С).<br /> (Copyright © 2016 Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1096-1194
Volume :
30
Issue :
5
Database :
MEDLINE
Journal :
Molecular and cellular probes
Publication Type :
Academic Journal
Accession number :
27720907
Full Text :
https://doi.org/10.1016/j.mcp.2016.10.003