Back to Search Start Over

Improving the off-axis spatial resolution and dynamic range of the NIF X-ray streak cameras (invited).

Authors :
MacPhee AG
Dymoke-Bradshaw AK
Hares JD
Hassett J
Hatch BW
Meadowcroft AL
Bell PM
Bradley DK
Datte PS
Landen OL
Palmer NE
Piston KW
Rekow VV
Hilsabeck TJ
Kilkenny JD
Source :
The Review of scientific instruments [Rev Sci Instrum] 2016 Nov; Vol. 87 (11), pp. 11E202.
Publication Year :
2016

Abstract

We report simulations and experiments that demonstrate an increase in spatial resolution of the NIF core diagnostic x-ray streak cameras by at least a factor of two, especially off axis. A design was achieved by using a corrector electron optic to flatten the field curvature at the detector plane and corroborated by measurement. In addition, particle in cell simulations were performed to identify the regions in the streak camera that contribute the most to space charge blurring. These simulations provide a tool for convolving synthetic pre-shot spectra with the instrument function so signal levels can be set to maximize dynamic range for the relevant part of the streak record.

Details

Language :
English
ISSN :
1089-7623
Volume :
87
Issue :
11
Database :
MEDLINE
Journal :
The Review of scientific instruments
Publication Type :
Academic Journal
Accession number :
27910532
Full Text :
https://doi.org/10.1063/1.4960376