Back to Search Start Over

First-in-man six-month results of a surface-modified coronary stent system in native coronary stenosis.

Authors :
Suwannasom P
Sotomi Y
Corti R
Kurz DJ
Roffi M
von Birgelen C
Buzzi S
Zucker A
Dijkstra J
Wykrzykowska JJ
de Winter RJ
Windecker S
Onuma Y
Serruys PW
Daemen J
Räber L
Source :
EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology [EuroIntervention] 2017 Apr 20; Vol. 12 (17), pp. 2118-2127.
Publication Year :
2017

Abstract

Aims: In preclinical studies, a bare metal cobalt-chromium stent with an active surface oxide layer modification (BMSmod) has been shown to inhibit neointimal hyperplasia effectively. We sought to assess both the clinical safety and feasibility of the BMSmod.<br />Methods and Results: In this prospective, non-randomised, first-in-man multicentre study, a total of 31 patients with de novo coronary lesions, reference lumen diameters of 2.5-3.5 mm and lesion length ≤16 mm, were enrolled. Quantitative coronary angiography and optical coherence tomography (OCT) were performed at baseline and six-month follow-up. Primary angiographic and OCT endpoints included in-stent late lumen loss (LLL) and mean neointimal thickness at six months. The device-oriented composite endpoint (DoCE), defined as cardiac death, myocardial infarction not clearly attributable to a non-intervention vessel, and clinically indicated target lesion revascularisation (CI-TLR), was analysed according to the intention-to-treat principle. In 31 patients (33 lesions), the procedural success rate was 93.5%. At six months, angiographic LLL was 0.91±0.45 mm and binary angiographic restenosis occurred in 23.3% of lesions. Out of 33 lesions, OCT was performed in 27 lesions at both time points. Mean neointimal thickness amounted to 348±116 µm. At six months, the DoCE was 19.4% due to the occurrence of CI-TLR in five patients (including one late definite stent thrombosis of a non-study stent).<br />Conclusions: In contrast to previous preclinical pathophysiological work, the BMSmod did not prevent neointimal hyperplasia in a first-in-man clinical setting.

Details

Language :
English
ISSN :
1969-6213
Volume :
12
Issue :
17
Database :
MEDLINE
Journal :
EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology
Publication Type :
Academic Journal
Accession number :
27993758
Full Text :