Back to Search
Start Over
The Reduction of Uncertainties for Absolute Piston Gage Pressure Measurements in the Atmospheric Pressure Range.
- Source :
-
Journal of research of the National Institute of Standards and Technology [J Res Natl Inst Stand Technol] 1989 Nov-Dec; Vol. 94 (6), pp. 343-346. - Publication Year :
- 1989
-
Abstract
- NIST pressure calibration services with nitrogen are now based on two transfer standard piston gages for which the effective areas have been determined by calibration with the manometer developed at NIST for gas thermometry. Root-sum-squared three sigma uncertainties for the areas for the two gages are 3.05 ppm and 4.18 ppm.
Details
- Language :
- English
- ISSN :
- 1044-677X
- Volume :
- 94
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- Journal of research of the National Institute of Standards and Technology
- Publication Type :
- Academic Journal
- Accession number :
- 28053420
- Full Text :
- https://doi.org/10.6028/jres.094.033