Back to Search
Start Over
Ten-year results of the Freedom Solo stentless heart valve: excellent haemodynamics but progressive valve dysfunction in the long term.
- Source :
-
Interactive cardiovascular and thoracic surgery [Interact Cardiovasc Thorac Surg] 2017 May 01; Vol. 24 (5), pp. 663-669. - Publication Year :
- 2017
-
Abstract
- Objectives: Freedom Solo (FS) is a pericardial stentless heart valve showing excellent haemodynamic performance at mid-term. The aim of this study was to evaluate the long-term performance of such bioprostheses.<br />Methods: Between December 2004 and November 2009, 109 patients (31 men; mean age 76 ± 6 years) underwent aortic valve replacement with FS. Preoperatively, the mean NYHA class was 2.5 ± 0.7, the mean EuroSCORE II, 2.8 ± 2.5. Mean prosthesis size was 22.7 ± 1.9 mm; concomitant procedures were performed in 65 patients. Structural valve deterioration (SVD) was diagnosed according to the Valve Academic Research Consortium-2 definition.<br />Results: Two patients (1.8%) died within 30 days. Follow-up (72 ± 36 months) was 100% completed. The 1-, 5- and 10-year actuarial survival rates were 89, 73 and 42%, respectively, with 8 valve-related deaths; the actuarial freedom from SVD was 99, 93 and 76%. During 61 ± 39 months of follow-up, echocardiographic findings worsened progressively: At discharge, 3-5 and 7-9 years, the mean gradient was 8 ± 4, 12 ± 11 and 19 ± 19 mmHg ( P < 0.01); the indexed effective orifice area was 1.0 ± 0.2, 0.9 ± 0.2 and 0.8 ± 0.3 cm 2 /m 2 ( P < 0.01). Of the 13 patients who developed SVD, it was due to aortic stenosis in 11. SVD was a predictor of cardiovascular mortality at univariate analysis (HR 2.87, 1.12-7.29); 2 explanted prostheses showed massive calcium deposits with mean calcium and phosphorus contents of 234 ± 16 and 116 ± 7 mg/g dry weight, respectively.<br />Conclusions: The FS bioprosthesis shows excellent mid-term clinical and haemodynamic results and offers an alternative to other valves, particularly in the case of a small aortic annulus. Worsening of FS performance was observed at late follow-up because of progressive SVD with stenosis, questioning whether it should be used in patients with a long life expectancy.<br /> (© The Author 2017. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.)
- Subjects :
- Aged
Aortic Valve diagnostic imaging
Aortic Valve Stenosis mortality
Aortic Valve Stenosis physiopathology
Echocardiography
Female
Follow-Up Studies
Humans
Italy epidemiology
Male
Prosthesis Design
Survival Rate trends
Treatment Outcome
Aortic Valve surgery
Aortic Valve Stenosis surgery
Bioprosthesis
Forecasting
Heart Valve Prosthesis Implantation methods
Hemodynamics physiology
Pericardium transplantation
Subjects
Details
- Language :
- English
- ISSN :
- 1569-9285
- Volume :
- 24
- Issue :
- 5
- Database :
- MEDLINE
- Journal :
- Interactive cardiovascular and thoracic surgery
- Publication Type :
- Academic Journal
- Accession number :
- 28329194
- Full Text :
- https://doi.org/10.1093/icvts/ivw391