Back to Search Start Over

Performance testing of the neutron flux monitors from 10keV to 1MeV developed for BNCT: A preliminary study.

Authors :
Guan X
Manabe M
Tamaki S
Sato F
Murata I
Wang T
Source :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2017 Jul; Vol. 125, pp. 119-123. Date of Electronic Publication: 2017 Apr 12.
Publication Year :
2017

Abstract

The neutron flux monitors from 10keV to 1MeV designed for boron neutron capture therapy (BNCT) were experimentally tested with prototype monitors in an appropriate neutron field produced at the intense deuterium-tritium neutron source facility OKTAVIAN of Osaka University, Japan. The experimental test results and related analysis indicated that the performance of the monitors was good and the neutron fluxes from 10keV to 1MeV of practical BNCT neutron sources can be measured within 10% by the monitors.<br /> (Copyright © 2017 Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1872-9800
Volume :
125
Database :
MEDLINE
Journal :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
Publication Type :
Academic Journal
Accession number :
28432929
Full Text :
https://doi.org/10.1016/j.apradiso.2017.04.014