Repeatability of topographic corneal thickness in keratoconus comparing Visante™ OCT and Oculus Pentacam HR ® topographer.

Authors :
Otchere H
Sorbara L
Source :
Contact lens & anterior eye : the journal of the British Contact Lens Association [Cont Lens Anterior Eye] 2017 Aug; Vol. 40 (4), pp. 217-223. Date of Electronic Publication: 2017 May 10.
Publication Year :
2017

Abstract

Purpose: To assess the repeatability and reproducibility of Visante™ OCT and Oculus Pentacam HR® in measuring central (CCT) and topographic corneal thickness (TCT) along the principle meridians in keratoconus (KC) corneae.<br />Method: Twenty participants diagnosed with KC were recruited. There were two study visits. On the first study visit, two repeated corneal thickness measurements were obtained with each instrument. Measurements were repeated at least 48hrs later in the same order. TCT were recorded in the 90, 180, 45 and 135 meridians at 1mm intervals across 8mm chord.<br />Results: Mean CCT for Visante™ OCT was 484.97±43.14μm (range: 484.84-486.09) and Oculus Pentacam HR® was 478.86±45.31μm (range: 477.20-480.53). No significant difference in TCT between the two visits (p=0.54) and measurements (p=0.63) for Visante™ OCT. For Oculus Pentacam HR®, no significant difference was found for each visit (p=0.18) but differences existed in the measurements outside of the central region (p=0.001). Tukey post-hoc analysis shows the differences (p˂0.05) were found in the +1 and +4 (supero-temporal) locations in the 135 meridian. Significant differences were found comparing the two instruments, (p<0.05). Bland Altman plots were used to demonstrate the differences between the two instruments and indicate their limits of agreement.<br />Conclusion: Both instruments gave repeatable measurements as no significant differences were found in most locations in all meridians. Comparing the two instruments, they were not reproducible in all locations.<br /> (Copyright © 2017 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1476-5411
Volume :
40
Issue :
4
Database :
MEDLINE
Journal :
Contact lens & anterior eye : the journal of the British Contact Lens Association
Publication Type :
Academic Journal
Accession number :
28501441
Full Text :
https://doi.org/10.1016/j.clae.2017.05.002