On-line Analysis of Catalytic Reaction Products Using a High-Pressure Tandem Micro-reactor GC/MS.

Authors :
Watanabe A
Kim YM
Hosaka A
Watanabe C
Teramae N
Ohtani H
Kim S
Park YK
Wang K
Freeman RR
Source :
Analytical sciences : the international journal of the Japan Society for Analytical Chemistry [Anal Sci] 2017; Vol. 33 (9), pp. 1085-1089.
Publication Year :
2017

Abstract

When a GC/MS system is coupled with a pressurized reactor, the separation efficiency and the retention time are directly affected by the reactor pressure. To keep the GC column flow rate constant irrespective of the reaction pressure, a restrictor capillary tube and an open split interface are attached between the GC injection port and the head of a GC separation column. The capability of the attached modules is demonstrated for the on-line GC/MS analysis of catalytic reaction products of a bio-oil model sample (guaiacol), produced under a pressure of 1 to 3 MPa.

Details

Language :
English
ISSN :
1348-2246
Volume :
33
Issue :
9
Database :
MEDLINE
Journal :
Analytical sciences : the international journal of the Japan Society for Analytical Chemistry
Publication Type :
Academic Journal
Accession number :
28890496
Full Text :
https://doi.org/10.2116/analsci.33.1085