Back to Search
Start Over
On-line Analysis of Catalytic Reaction Products Using a High-Pressure Tandem Micro-reactor GC/MS.
- Source :
-
Analytical sciences : the international journal of the Japan Society for Analytical Chemistry [Anal Sci] 2017; Vol. 33 (9), pp. 1085-1089. - Publication Year :
- 2017
-
Abstract
- When a GC/MS system is coupled with a pressurized reactor, the separation efficiency and the retention time are directly affected by the reactor pressure. To keep the GC column flow rate constant irrespective of the reaction pressure, a restrictor capillary tube and an open split interface are attached between the GC injection port and the head of a GC separation column. The capability of the attached modules is demonstrated for the on-line GC/MS analysis of catalytic reaction products of a bio-oil model sample (guaiacol), produced under a pressure of 1 to 3 MPa.
Details
- Language :
- English
- ISSN :
- 1348-2246
- Volume :
- 33
- Issue :
- 9
- Database :
- MEDLINE
- Journal :
- Analytical sciences : the international journal of the Japan Society for Analytical Chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 28890496
- Full Text :
- https://doi.org/10.2116/analsci.33.1085