Back to Search Start Over

In Vitro Comparison of Two Barbed Suture Configurations for Flexor Tendon Repair.

Authors :
Engel JL
Gabra JN
Esterle AR
Lanzinger WD
Elias JJ
Source :
Journal of hand and microsurgery [J Hand Microsurg] 2017 Dec; Vol. 9 (3), pp. 126-130. Date of Electronic Publication: 2017 Nov 06.
Publication Year :
2017

Abstract

Purpose  The current study compares the strength of a previously studied technique for flexor tendon repair with barbed sutures to an experimental approach that aligns all the barbs to oppose distraction. Methods  Twelve flexor tendons from cadaveric specimens were mechanically tested following repair of simulated zone II tendon injuries. Two repair techniques utilizing barbed sutures were studied: the Marrero-Amadeo four-core barbed suture approach and the experimental configuration with all barbs on four cores opposing distraction. Maximum applied load at failure, that is, ultimate load, and 2-mm gap force were compared between the two repair techniques, both as raw values and after normalization to cross-sectional area of the intact tendon. Statistical testing was performed using t -tests and Mann-Whitney U -tests, where appropriate, with a significance level of 0.05. Results  The ultimate loads, raw (58.2 N) and normalized (4.8 N/mm <superscript>2</superscript> ), were significantly larger for the Marrero-Amadeo repair than the proposed experimental approach (35.6 N and 2.7 N/mm <superscript>2</superscript> ; p  < 0.05). No significant differences were identified for 2-mm gap force. Conclusion  The innovation of suturing to align all barbs to oppose distraction does not improve strength of the repaired tendon. The Marrero-Amadeo technique was found to have superior strength for use in traumatic zone II flexor tendon injuries.

Details

Language :
English
ISSN :
0974-3227
Volume :
9
Issue :
3
Database :
MEDLINE
Journal :
Journal of hand and microsurgery
Publication Type :
Academic Journal
Accession number :
29302136
Full Text :
https://doi.org/10.1055/s-0037-1606628