In Vitro Quantification of Gutter Formation and Chimney Graft Compression in Chimney EVAR Stent-Graft Configurations Using Electrocardiography-Gated Computed Tomography.

Authors :
Overeem SP
Donselaar EJ
Boersen JT
Groot Jebbink E
Slump CH
de Vries JPM
Reijnen MMPJ
Source :
Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists [J Endovasc Ther] 2018 Jun; Vol. 25 (3), pp. 387-394. Date of Electronic Publication: 2018 Mar 16.
Publication Year :
2018

Abstract

Purpose: To assess the dynamic behavior of chimney grafts during the cardiac cycle.<br />Methods: Three chimney endovascular aneurysm repair (EVAR) stent-graft configurations (Endurant and Advanta V12, Endurant and Viabahn, and Endurant and BeGraft) were placed in silicone aneurysm models and subjected to physiologic flow. Electrocardiography (ECG)-gated contrast-enhanced computed tomography was used to visualize geometric changes during the cardiac cycle. Endograft and chimney graft surface, gutter volume, chimney graft angulation over the center lumen line, and the D-ratio (the ratio between the lengths of the major and minor axes) were independently assessed by 2 observers at 10 time points in the cardiac cycle.<br />Results: Both gutter volumes and chimney graft geometry changed significantly during the cardiac cycle in all 3 configurations (p<0.001). Gutters and endoleaks were observed in all configurations. The largest gutter volume (232.8 mm <superscript>3</superscript> ) and change in volume (20.7 mm <superscript>3</superscript> ) between systole and diastole were observed in the Endurant-Advanta configuration. These values were 2.7- and 3.0-fold higher, respectively, compared to the Endurant-Viabahn configuration and 1.7- and 1.6-fold higher as observed in the Endurant-BeGraft configuration. The Endurant-Viabahn configuration had the highest D-ratio (right, 1.26-1.35; left, 1.33-1.48), while the Endurant-BeGraft configuration had the lowest (right, 1.11-1.17; left, 1.08-1.15). Assessment of the interobserver variability showed a high correlation (intraclass correlation >0.935) between measurements.<br />Conclusion: Gutter volumes and stent compression are dynamic phenomena that reshape during the cardiac cycle. Compelling differences were observed during the cardiac cycle in all configurations, with the self-expanding (Endurant-Viabahn) chimney EVAR configurations having smaller gutters and less variation in gutter volume during the cardiac cycle yet more stent compression without affecting the chimney graft surface.

Details

Language :
English
ISSN :
1545-1550
Volume :
25
Issue :
3
Database :
MEDLINE
Journal :
Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists
Publication Type :
Academic Journal
Accession number :
29544371
Full Text :
https://doi.org/10.1177/1526602818762399