In Vitro Quantification of Gutter Formation and Chimney Graft Compression in Chimney EVAR Stent-Graft Configurations Using Electrocardiography-Gated Computed Tomography.
- Source :
-
Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists [J Endovasc Ther] 2018 Jun; Vol. 25 (3), pp. 387-394. Date of Electronic Publication: 2018 Mar 16. - Publication Year :
- 2018
-
Abstract
- Purpose: To assess the dynamic behavior of chimney grafts during the cardiac cycle.<br />Methods: Three chimney endovascular aneurysm repair (EVAR) stent-graft configurations (Endurant and Advanta V12, Endurant and Viabahn, and Endurant and BeGraft) were placed in silicone aneurysm models and subjected to physiologic flow. Electrocardiography (ECG)-gated contrast-enhanced computed tomography was used to visualize geometric changes during the cardiac cycle. Endograft and chimney graft surface, gutter volume, chimney graft angulation over the center lumen line, and the D-ratio (the ratio between the lengths of the major and minor axes) were independently assessed by 2 observers at 10 time points in the cardiac cycle.<br />Results: Both gutter volumes and chimney graft geometry changed significantly during the cardiac cycle in all 3 configurations (p<0.001). Gutters and endoleaks were observed in all configurations. The largest gutter volume (232.8 mm <superscript>3</superscript> ) and change in volume (20.7 mm <superscript>3</superscript> ) between systole and diastole were observed in the Endurant-Advanta configuration. These values were 2.7- and 3.0-fold higher, respectively, compared to the Endurant-Viabahn configuration and 1.7- and 1.6-fold higher as observed in the Endurant-BeGraft configuration. The Endurant-Viabahn configuration had the highest D-ratio (right, 1.26-1.35; left, 1.33-1.48), while the Endurant-BeGraft configuration had the lowest (right, 1.11-1.17; left, 1.08-1.15). Assessment of the interobserver variability showed a high correlation (intraclass correlation >0.935) between measurements.<br />Conclusion: Gutter volumes and stent compression are dynamic phenomena that reshape during the cardiac cycle. Compelling differences were observed during the cardiac cycle in all configurations, with the self-expanding (Endurant-Viabahn) chimney EVAR configurations having smaller gutters and less variation in gutter volume during the cardiac cycle yet more stent compression without affecting the chimney graft surface.
- Subjects :
- Aortic Aneurysm, Abdominal diagnostic imaging
Aortic Aneurysm, Abdominal physiopathology
Arterial Pressure
Materials Testing
Models, Anatomic
Models, Cardiovascular
Prosthesis Design
Pulsatile Flow
Aortic Aneurysm, Abdominal surgery
Blood Vessel Prosthesis
Blood Vessel Prosthesis Implantation instrumentation
Cardiac-Gated Imaging Techniques
Electrocardiography
Endovascular Procedures instrumentation
Multidetector Computed Tomography
Stents
Subjects
Details
- Language :
- English
- ISSN :
- 1545-1550
- Volume :
- 25
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists
- Publication Type :
- Academic Journal
- Accession number :
- 29544371
- Full Text :
- https://doi.org/10.1177/1526602818762399