Back to Search Start Over

Quantification of metronidazole in human plasma using a highly sensitive and rugged LC-MS/MS method for a bioequivalence study.

Authors :
Vanol PG
Sanyal M
Shah PA
Shrivastav PS
Source :
Biomedical chromatography : BMC [Biomed Chromatogr] 2018 Aug; Vol. 32 (8), pp. e4242. Date of Electronic Publication: 2018 Apr 22.
Publication Year :
2018

Abstract

A highly sensitive, selective and rugged method has been described for the quantification of metronidazole (MTZ) in human plasma by liquid chromatography-tandem mass spectrometry using metronidazole-d4 as the internal standard (IS). The analyte and the IS were extracted from 100 μL plasma by liquid-liquid extraction. The clear samples obtained were chromatographed on an ACE C <subscript>18</subscript> (100 × 4.6 mm, 5 μm) column using acetonitrile and 10.0 mm ammonium formate in water, pH 4.00 (80:20, v/v) as the mobile phase. A triple quadrupole mass spectrometer system equipped with turbo ion spray source and operated in multiple reaction monitoring mode was used for the detection and quantification of MTZ. The calibration range was established from 0.01 to 10.0 μg/mL. The results of validation testing for precision and accuracy, selectivity, matrix effects, recovery and stability complied with current bioanalytical guidelines. A run time of 3.0 min permitted analysis of more than 300 samples in a day. The method was applied to a bioequivalence study with 250 mg MTZ tablet formulation in 24 healthy Indian males.<br /> (Copyright © 2018 John Wiley & Sons, Ltd.)

Details

Language :
English
ISSN :
1099-0801
Volume :
32
Issue :
8
Database :
MEDLINE
Journal :
Biomedical chromatography : BMC
Publication Type :
Academic Journal
Accession number :
29572903
Full Text :
https://doi.org/10.1002/bmc.4242