Evaluation of the Vitek MS v3.0 Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry System for Identification of Mycobacterium and Nocardia Species.

Authors :
Body BA
Beard MA
Slechta ES
Hanson KE
Barker AP
Babady NE
McMillen T
Tang YW
Brown-Elliott BA
Iakhiaeva E
Vasireddy R
Vasireddy S
Smith T
Wallace RJ Jr
Turner S
Curtis L
Butler-Wu S
Rychert J
Source :
Journal of clinical microbiology [J Clin Microbiol] 2018 May 25; Vol. 56 (6). Date of Electronic Publication: 2018 May 25 (Print Publication: 2018).
Publication Year :
2018

Abstract

This multicenter study was designed to assess the accuracy and reproducibility of the Vitek MS v3.0 matrix-assisted laser desorption ionization-time of flight (MALDI-TOF) mass spectrometry system for identification of Mycobacterium and Nocardia species compared to DNA sequencing. A total of 963 clinical isolates representing 51 taxa were evaluated. In all, 663 isolates were correctly identified to the species level (69%), with another 231 (24%) correctly identified to the complex or group level. Fifty-five isolates (6%) could not be identified despite repeat testing. All of the tuberculous mycobacteria (45/45; 100%) and most of the nontuberculous mycobacteria (569/606; 94%) were correctly identified at least to the group or complex level. However, not all species or subspecies within the M. tuberculosis , M. abscessus , and M. avium complexes and within the M. fortuitum and M. mucogenicum groups could be differentiated. Among the 312 Nocardia isolates tested, 236 (76%) were correctly identified to the species level, with an additional 44 (14%) correctly identified to the complex level. Species within the N. nova and N. transvalensis complexes could not always be differentiated. Eleven percent of the isolates (103/963) underwent repeat testing in order to get a final result. Identification of a representative set of Mycobacterium and Nocardia species was highly reproducible, with 297 of 300 (99%) replicates correctly identified using multiple kit lots, instruments, analysts, and sites. These findings demonstrate that the system is robust and has utility for the routine identification of mycobacteria and Nocardia in clinical practice.<br /> (Copyright © 2018 American Society for Microbiology.)

Details

Language :
English
ISSN :
1098-660X
Volume :
56
Issue :
6
Database :
MEDLINE
Journal :
Journal of clinical microbiology
Publication Type :
Academic Journal
Accession number :
29643203
Full Text :
https://doi.org/10.1128/JCM.00237-18