Back to Search Start Over

Analytical characteristics of nano-electrospray operated under super-atmospheric pressure.

Authors :
Rahman MM
Chen LC
Source :
Analytica chimica acta [Anal Chim Acta] 2018 Aug 27; Vol. 1021, pp. 78-84. Date of Electronic Publication: 2018 Mar 29.
Publication Year :
2018

Abstract

High-pressure nanoelectrospray ionization (nanoESI) source is a recently developed technique in which the electrospray ionization is generated inside an enclosed chamber with gas pressure higher than the atmospheric pressure. In this paper, the performance of nanoESI under different gas pressures, emitter position, ion inlet temperature, additive for desalination are presented. Under a pressure of 2 bars, the nanoESI is almost eased from the electrical discharge problem, and that offers a wider tuning window for the emitter potential to produces a higher and more stable ion signal. With optimized ion inlet temperature, the high-pressure operation facilitates the generation of ion species of higher charge-state from the highly aqueous solution, and produced less sodium adducts. A preparation method for the high-throughput analysis of raw biological samples using disposable plastic nanoESI emitter is also described.<br /> (Copyright © 2018 Elsevier B.V. All rights reserved.)

Details