Atrioventricular Interval Extension Is Highly Efficient in Preventing Unnecessary Right Ventricular Pacing in Sinus Node Disease: A Randomized Cross-Over Study Versus Dual- to Atrial Single-Chamber Mode Switch.

Authors :
Calvi V
Pisanò EC
Brieda M
Melissano D
Castaldi B
Guastaferro C
Nigro G
Madalosso M
Orsida D
Rovai N
Gargaro A
Capucci A
Source :
JACC. Clinical electrophysiology [JACC Clin Electrophysiol] 2017 May; Vol. 3 (5), pp. 482-490. Date of Electronic Publication: 2017 Feb 01.
Publication Year :
2017

Abstract

Objectives: This study sought to compare the Intrinsic Rhythm Support (IRSplus) and Ventricular Pace Suppress (VpS) in terms of right ventricular pacing percentage (VP %), mean atrioventricular interval (MAVI), atrial fibrillation, and cardiac volumes.<br />Background: Modern pacemakers are provided with algorithms for reducing unnecessary ventricular pacing. These may be classified as: periodic search for intrinsic atrioventricular (AV) conduction prolonging the AV delay accordingly; or DDD-ADI mode switch. The IRSplus and VpS algorithms belong to the former and latter classes, respectively.<br />Methods: Patients with sick sinus dysfunction without evidence of II/III degree AV block were 1:1 randomized to 6-month periods of either IRSplus or VpS, and then crossed over. Subsequent follow-ups were at the 12th month after randomization for device data retrieving, and at the 18th month with the same device programming for echocardiographic assessment.<br />Results: A total of 230 patients (62% males, median age 75 years [interquartile range: 69 to 79 years]) were enrolled. At a linear mixed-model analysis with order of treatment and investigational sites as nested random effects, differences in VP% and MAVI reached statistical significance: VP% was 1% (0% to 11%) during IRSplus and 3% (0% to 26%) during VpS (p = 0.029); MAVI was 225 ms (198 to 253 ms) during IRSplus and 214 ms (188 to 240 ms) during VpS (p = 0.014). No differences were observed in atrial fibrillation burden and incidence, ejection fraction, and cardiac volumes.<br />Conclusions: Both IRSplus and VpS algorithms ensured VP% ≤3% in most patients with sinus node dysfunction and preserved AV conduction. The IRSplus was slightly more efficient in reducing VP% at the expense of a small MAVI increase, with statistical but clinically insignificant differences. (Ventricular Pace Suppression Versus Intrinsic Rhythm Support Study; NCT01528657).<br /> (Copyright © 2017 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.)

Details

Language :
English
ISSN :
2405-5018
Volume :
3
Issue :
5
Database :
MEDLINE
Journal :
JACC. Clinical electrophysiology
Publication Type :
Academic Journal
Accession number :
29759604
Full Text :
https://doi.org/10.1016/j.jacep.2016.11.011