Back to Search
Start Over
Metal-on-metal surface hip arthroplasties. Is annual monitoring of blood metal levels necessary?
- Source :
-
Revista espanola de cirugia ortopedica y traumatologia (English ed.) [Rev Esp Cir Ortop Traumatol (Engl Ed)] 2018 Nov - Dec; Vol. 62 (6), pp. 436-441. Date of Electronic Publication: 2018 Jun 06. - Publication Year :
- 2018
-
Abstract
- Introduction: Blood cobalt (Co) and chromium (Cr) ion levels have been used as surveillance tools for adverse reaction to metal debris in metal-on-metal (MoM) hip arthroplasty. The aim of our study was to present serial 7-13 year blood Co and Cr levels in a cohort of MoM total hip arthroplasties in asymptomatic patients.<br />Material and Methods: A total of 52 MoM surface total hip arthroplasties were included in this study with data collected prospectively. Annual follow-up with blood Co and Cr measurements was performed. Revision surgery was necessary for 16 patients and therefore they dropped out of the study. The metal ion levels were analyzed separately in 31 asymptomatic patients.<br />Results: High Cr or Co levels were not found continuously in any of the asymptomatic patients. The median Cr in blood was maintained between 1.3 and 5.4μg/L and that of Co between 0.5 and 1.2μg/L. After 7 years, there was no significant change in Co and Cr values.<br />Conclusion: In the medium term and in asymptomatic patients, the metallic blood levels remained unchanged; therefore, it does not seem appropriate to repeat these tests on a permanent basis in annual controls.<br /> (Copyright © 2018 SECOT. Publicado por Elsevier España, S.L.U. All rights reserved.)
Details
- Language :
- English; Spanish; Castilian
- ISSN :
- 2173-576X
- Volume :
- 62
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- Revista espanola de cirugia ortopedica y traumatologia (English ed.)
- Publication Type :
- Academic Journal
- Accession number :
- 29884514
- Full Text :
- https://doi.org/10.1016/j.recot.2018.03.001