Back to Search Start Over

Spinal Surgery With Electrically Evoked Potential Monitoring and Monopolar Electrocautery: Is Prior Removal of a Cochlear Implant Necessary?

Authors :
Studer D
Stieger C
Reichlin CJ
Terrier A
Allum JHJ
Source :
Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology [Otol Neurotol] 2019 Jan; Vol. 40 (1), pp. e7-e13.
Publication Year :
2019

Abstract

: Transcranial electric stimulation to generate motor evoked potentials in lower limb muscles is the standard technique used to monitor spinal cord efferent pathways during surgical correction for spinal deformities. Monopolar electrical cauterization is also used by default in the thoracic and lumbar area of the spine during this kind of surgery to prevent major blood loss. Owing to the high levels of current used, both techniques are considered contraindicative if the patient has a cochlear implant (CI). Here, we present a CI patient who underwent corrective spinal fusion surgery for a severe kyphoscoliotic spinal deformity on whom both techniques were used without any negative effects on the CI function. A major improvement in sagittal body balance was achieved with no loss in implant-aided hearing levels. These results add to reports that CI manufactures should review their evidence underlying recommendations that transcranial electric stimulation and upper thoracic monopolar electrical cauterization are high risk for CI users, possibly initiating verification studies.

Details

Language :
English
ISSN :
1537-4505
Volume :
40
Issue :
1
Database :
MEDLINE
Journal :
Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology
Publication Type :
Academic Journal
Accession number :
30371634
Full Text :
https://doi.org/10.1097/MAO.0000000000002062