Accuracy and precision of proton density fat fraction measurement across field strengths and scan intervals: A phantom and human study.

Authors :
Kim HJ
Cho HJ
Kim B
You MW
Lee JH
Huh J
Kim JK
Source :
Journal of magnetic resonance imaging : JMRI [J Magn Reson Imaging] 2019 Jul; Vol. 50 (1), pp. 305-314. Date of Electronic Publication: 2018 Nov 14.
Publication Year :
2019

Abstract

Background: Complex-based chemical shift imaging-based magnetic resonance imaging (CSE-MRI) is emerging as a preferred method for noninvasively quantifying proton density fat fraction (PDFF), a promising quantitative imaging biomarker (QIB) for longitudinal hepatic steatosis measurement.<br />Purpose: To determine linearity, bias, repeatability, and reproducibility of the PDFF measurement using CSE-MRI (CSE-PDFF) across scan intervals, MR field strengths, and readers in phantom and nonalcoholic fatty liver disease (NAFLD) patients.<br />Study Type: Institutional Review Board (IRB)-approved prospective.<br />Subjects: Fat-water phantom and 20 adult patients.<br />Field Strength/sequence: 1.5 T and 3.0 T MR systems and a commercially available CSE-MRI sequence (IDEAL-IQ).<br />Assessment: Two independent readers measured CSE-PDFF of fat-water phantom and NAFLD patients across two field strengths and scan intervals (same-day and 2-week) each and in a combination of both. MR spectroscopy-based PDFF (MRS-PDFF) was used as the reference standard for phantom PDFF.<br />Statistical Tests: Linearity and bias of measurement were evaluated by linear regression analysis and Bland-Altman plots, respectively. Repeatability and reproducibility were assessed by coefficient of variance and repeatability / reproducibility coefficients (RC). The intraclass correlation coefficient was used to validate intra- and interobserver agreements.<br />Results: CSE-PDFF showed high linearity and small bias (-0.6-0.4 PDFF%) with 95% limits of agreement within ±2.9 PDFF% across field strengths, 2-week interscan period, and readers in the clinical scans. CSE-PDFF was highly repeatable and reproducible both in phantom and clinical scans, with the largest observed RC across field strengths and 2-week interscan period being 3 PDFF%.<br />Data Conclusion: CSE-PDFF is a robust QIB with high linearity, small bias, and excellent repeatability/reproducibility. A change of more than 3 PDFF% across field strengths within 2 weeks of scan interval likely reflects a true change, which is well within the clinically acceptable range.<br />Level of Evidence: 3 Technical Efficacy: Stage 2 J. Magn. Reson. Imaging 2019;50:305-314.<br /> (© 2018 International Society for Magnetic Resonance in Medicine.)

Details

Language :
English
ISSN :
1522-2586
Volume :
50
Issue :
1
Database :
MEDLINE
Journal :
Journal of magnetic resonance imaging : JMRI
Publication Type :
Academic Journal
Accession number :
30430684
Full Text :
https://doi.org/10.1002/jmri.26575