COMPARISON OF ORGAN-BASED TUBE CURRENT MODULATION AND BISMUTH SHIELDING IN CHEST CT: EFFECT ON THE IMAGE QUALITY AND THE PATIENT DOSE.
- Source :
-
Radiation protection dosimetry [Radiat Prot Dosimetry] 2019 Nov 30; Vol. 185 (1), pp. 42-48. - Publication Year :
- 2019
-
Abstract
- The aim of the study was to compare the absorbed doses and image quality of organ-based tube current modulation (OBTCM) and bismuth shielding of breasts and thyroid against regular tube current modulation in chest CT scan. An anthropomorphic phantom and MOSFET dosemeters were used to evaluate absorbed doses. Image quality was assessed from HU and noise. Relative to the reference scan, the average absorbed dose reduction with OBTCM was 5.2% and with bismuth shields 24.2%. Difference in HU values compared to the reference varied between -4.1 and 4.2 HU in OBTCM scan and between -22.2 and 118.6 HU with bismuth shields. Image noise levels varied between 10.0 to 26.3 HU in the reference scan, from 9.6 to 27.7 HU for the OBTCM scan and from 11.9 to 43.9 HU in the bismuth scan. The use of bismuth shields provided greatest dose reduction compared to the investigated OBTCM.<br /> (© The Author(s) 2019. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.)
- Subjects :
- Breast radiation effects
Female
Humans
Organ Specificity
Protective Devices
Radiation Dosage
Radiometry methods
Thorax radiation effects
Thyroid Gland radiation effects
Tomography, X-Ray Computed methods
Bismuth chemistry
Image Processing, Computer-Assisted methods
Phantoms, Imaging
Radiation Protection instrumentation
Radiography, Thoracic standards
Tomography, X-Ray Computed standards
Subjects
Details
- Language :
- English
- ISSN :
- 1742-3406
- Volume :
- 185
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Radiation protection dosimetry
- Publication Type :
- Academic Journal
- Accession number :
- 30544171
- Full Text :
- https://doi.org/10.1093/rpd/ncy242