Rotational position of the tibial component can decrease bony coverage of the tibial component in Oxford mobile-bearing unicompartmental knee arthroplasty.
- Source :
-
The Knee [Knee] 2019 Mar; Vol. 26 (2), pp. 459-465. Date of Electronic Publication: 2019 Jan 27. - Publication Year :
- 2019
-
Abstract
- Background: This study examined how coverage of the tibial component changes when the tibia vertical cut is externally or internally rotated in Oxford mobile-bearing unicompartmental knee arthroplasty.<br />Materials and Methods: Fifty patients scheduled for primary Oxford medial unicompartmental knee arthroplasty (UKA) at the current hospital were included in this study. This study was a computed tomography (CT) simulation study. The anteroposterior (AP) and mediolateral (ML) length as well as the ML/AP ratio of the tibial cut surfaces were calculated when the vertical cut was performed parallel (base line), five degrees externally rotated (ER5), 10° externally rotated (ER10), five degrees internally rotated (IR5), or 10° internally rotated (IR10) relative to the tibial AP line using pre-operative CT. The tibial AP line connecting the middle of the posterior collateral ligament to the medial border of the patellar tendon attachment is a reproducible and reliable line because it is perpendicular to the SEA. These parameters among three lines were compared using a repeated measures ANOVA.<br />Results: The mean ML/AP ratios were statistically significantly lower in ER5 (0.53 ± 0.04) than base line (0.56 ± 0.04) (P < 0.01). The ER10 (0.48 ± 0.03) also exhibited lower mean ML/AP ratios than ER5 (0.53 ± 0.04) (P < 0.01). The mean ML/AP ratios were higher in IR5 (0.59 ± 0.04) than base line (0.56 ± 0.04) (P < 0.01). The IR10 (0.63 ± 0.06) also showed a higher mean ML/AP ratio than IR5 (0.59 ± 0.04) (P < 0.01).<br />Conclusion: Rotational malalignment of tibial vertical cuts can affect tibial coverage in Oxford mobile-bearing unicompartmental knee arthroplasty.<br /> (Copyright © 2019 Elsevier B.V. All rights reserved.)
- Subjects :
- Aged
Arthroplasty, Replacement, Knee instrumentation
Arthroplasty, Replacement, Knee methods
Bone Malalignment etiology
Computer Simulation
Female
Humans
Male
Middle Aged
Tomography, X-Ray Computed
Arthroplasty, Replacement, Knee adverse effects
Knee Joint surgery
Knee Prosthesis adverse effects
Osteoarthritis, Knee surgery
Rotation adverse effects
Tibia surgery
Subjects
Details
- Language :
- English
- ISSN :
- 1873-5800
- Volume :
- 26
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- The Knee
- Publication Type :
- Academic Journal
- Accession number :
- 30700389
- Full Text :
- https://doi.org/10.1016/j.knee.2019.01.003