Back to Search
Start Over
Relative versus absolute rises in T/QRS ratio by ST analysis of fetal electrocardiograms in labour: A case-control pilot study.
- Source :
-
PloS one [PLoS One] 2019 Mar 26; Vol. 14 (3), pp. e0214357. Date of Electronic Publication: 2019 Mar 26 (Print Publication: 2019). - Publication Year :
- 2019
-
Abstract
- Introduction: The additional value of ST analysis during labour is uncertain. In ST analysis, a T/QRS baseline value is calculated from the fetal electrocardiogram and successive T/QRS ratios are compared to this baseline. However, variation in the orientation of the electrical heart axis between fetuses may yield different T/QRS baseline values. In case of a higher T/QRS baseline value more ST events are encountered, although not always related to perinatal outcome. We hypothesised that we can partly correct for this effect by analysing T/QRS rises as a percentage from baseline (relative ST analysis). This study aimed to explore whether relative ST analysis has better diagnostic value for cord acidaemia compared to conventional ST analysis, where predefined fixed T/QRS ratios are used.<br />Methods and Materials: A case-control study was performed in 20 term human fetuses during labour; 10 cases (umbilical cord artery pH <7.05 at birth, defining acidaemia) and 10 controls (pH >7.20) were included. The fetal electrocardiogram was recorded using a STAN monitor. We electronically extracted all T/QRS values, baseline and episodic ST events from the STAN monitor and calculated the relative T/QRS changes. The cut-off for relative ST events was determined in a receiver operator characteristic (ROC) curve at optimal specificity for cord acidaemia. Parameters of interest were area under the curve (AUC) of the ROC curve for relative ST events and test performance of both conventional and relative ST analysis.<br />Results: Relative ST analysis showed an AUC of 0.99. The optimal cut-off value for relative T/QRS rise was determined at 0.70. Relative vs conventional (absolute) ST analysis showed a specificity of 100% vs 40% (p = 0.031); sensitivity 90% vs 90%; positive likelihood ratio infinity vs 1.5; negative likelihood ratio 0.10 vs 0.25, respectively.<br />Conclusion: Relative ST analysis seems to be a promising method to detect impending fetal acidaemia during labour. Further studies are required to determine the diagnostic accuracy.<br />Competing Interests: Guid Oei is head of the research group "fundamental perinatology", which is a collaboration of the Eindhoven University of Technology and the Máxima Medical Center. He was involved in the founding of Nemo Healthcare BV, that arised from this collaboration. Guid Oei is not a shareholder in this company. Rik Vullings is shareholder in Nemo Healthcare BV, the Netherlands. None of the other authors have any conflicts of interest. This does not alter our adherence to PLOS ONE policies on sharing data and materials.
Details
- Language :
- English
- ISSN :
- 1932-6203
- Volume :
- 14
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- PloS one
- Publication Type :
- Academic Journal
- Accession number :
- 30913253
- Full Text :
- https://doi.org/10.1371/journal.pone.0214357