- Source :
-
Analytical chemistry [Anal Chem] 2019 Jun 18; Vol. 91 (12), pp. 7510-7515. Date of Electronic Publication: 2019 Jun 04. - Publication Year :
- 2019
-
Abstract
- The introduction of more sensitive mass spectrometers allows researchers to adapt front-end liquid chromatography (LC) to individual needs for the analysis of complex proteomes. Where absolute sensitivity is not paramount, it is advantageous to switch from a highly sensitive nanoflow-LC setup, the de facto standard platform in mass-spectrometry (MS)-based discovery proteomics, to a more robust, high-throughput-compatible microflow or conventional-flow setup. To enhance the microflow-LC-MS electrospray process of complex proteomic samples, we tested the effects of different solvents, including 2-propanol, methanol, and acetonitrile, pure or as mixture with dimethyl sulfoxide, which were added postcolumn to the eluting sample. Postcolumn addition of organic solvents strongly enhanced the electrospray efficiency in microflow-LC-MS experiments and improved the sensitivity across the entire gradient and for early eluting peptides by up to 10-fold. Postcolumn solvent addition did not negatively affect chromatographic performance and resulted in an overall 28-36% increase in identifications at both the protein and peptide levels. The presented microflow-LC-MS workflow, including postcolumn solvent addition, can be easily adopted on any LC-MS/MS platform.
Details
- Language :
- English
- ISSN :
- 1520-6882
- Volume :
- 91
- Issue :
- 12
- Database :
- MEDLINE
- Journal :
- Analytical chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 31117400
- Full Text :
- https://doi.org/10.1021/acs.analchem.9b00118