Back to Search
Start Over
Final Two-Year Outcomes for the Sentry Bioconvertible Inferior Vena Cava Filter in Patients Requiring Temporary Protection from Pulmonary Embolism.
- Source :
-
Journal of vascular and interventional radiology : JVIR [J Vasc Interv Radiol] 2020 Feb; Vol. 31 (2), pp. 221-230.e3. Date of Electronic Publication: 2019 Nov 08. - Publication Year :
- 2020
-
Abstract
- Purpose: To report final 2-year outcomes with the Sentry bioconvertible inferior vena cava (IVC) filter in patients requiring temporary protection against pulmonary embolism (PE).<br />Materials and Methods: In a prospective multicenter trial, the Sentry filter was implanted in 129 patients with documented deep vein thrombosis (DVT) and/or PE (67.5%) or who were at temporary risk of developing DVT/PE (32.6%). Patients were monitored and bioconversion status ascertained by radiography, computed tomography (CT), and CT venography through 2 years.<br />Results: The composite primary 6-month endpoint of clinical success was achieved in 97.4% (111/114) of patients. The rate of new symptomatic PE was 0% (n = 126) through 1 year and 2.4% (n = 85) through the second year of follow-up, with 2 new nonfatal cases at 581 and 624 days that were adjudicated as not related to the procedure or device. Two patients (1.6%) developed symptomatic caval thrombosis during the first month and underwent successful interventions without recurrence. No other filter-related symptomatic complications occurred through 2 years. There was no filter tilting, migration, embolization, fracture, or caval perforation and no filter-related deaths through 2 years. Filter bioconversion was successful for 95.7% (110/115) of patients at 6 months, 96.4% (106/110) of patients at 12 months, and 96.5% (82/85) of patients at 24 months. Through 24 months of follow-up, there was no evidence of late-stage IVC obstruction or thrombosis after filter bioconversion or of thrombogenicity associated with retracted filter arms.<br />Conclusions: The Sentry IVC filter provided safe and effective protection against PE, with a high rate of intended bioconversion and a low rate of device-related complications, through 2 years of follow-up.<br /> (Copyright © 2019 SIR. Published by Elsevier Inc. All rights reserved.)
- Subjects :
- Adult
Aged
Aged, 80 and over
Belgium
Chile
Computed Tomography Angiography
Female
Humans
Male
Middle Aged
Phlebography
Prosthesis Design
Prosthesis Implantation adverse effects
Pulmonary Embolism diagnostic imaging
Risk Factors
Time Factors
Treatment Outcome
United States
Venous Thrombosis diagnostic imaging
Young Adult
Prosthesis Implantation instrumentation
Pulmonary Embolism prevention & control
Vena Cava Filters
Vena Cava, Inferior diagnostic imaging
Venous Thrombosis prevention & control
Subjects
Details
- Language :
- English
- ISSN :
- 1535-7732
- Volume :
- 31
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- Journal of vascular and interventional radiology : JVIR
- Publication Type :
- Academic Journal
- Accession number :
- 31711748
- Full Text :
- https://doi.org/10.1016/j.jvir.2019.08.036