THERMAL AND FAST NEUTRON DOSE EQUIVALENT DISTRIBUTION MEASUREMENT OF 15-MV LINEAR ACCELERATOR USING A CR-39 NUCLEAR TRACK DETECTORS.

Authors :
Mohammadi S
Behmadi M
Mohammadi A
Toossi MTB
Source :
Radiation protection dosimetry [Radiat Prot Dosimetry] 2020 Jul 02; Vol. 188 (4), pp. 503-507.
Publication Year :
2020

Abstract

The main purpose of this study is to measure the contribution of the thermal and fast neutron dose along the central axis of the 15 MV Elekta Precise linac in a tissue equivalent phantom. In order to achieve this purpose, different points were selected in three field sizes of 5 × 5 cm2, 10 × 10 cm2 and 15 × 15 cm2. Fast and thermal neutrons were measured using CR-39 nuclear track detectors with and without thermal neutron converter of 10B, respectively. According to the results, the fast neutron dose equivalent was decreased as the depth increased (field size 5 × 5, 10 × 10 and 15 × 15 cm2 fall from 0.35 to 0.15, 0.5 to 0.3 and 0.5 to 0.3, respectively). Thermal dose equivalent was increased as the depth increased in the tissue equivalent phantom (field size 5 × 5, 10 × 10 and 15 × 15 cm2 rise from 0.1 to 0.4, 0.4 to 0.8 and 0.4 to 0.9, respectively). In conclusion, at depth <3 cm, most existing neutrons are fast and CR-39 films are sensitive to fast neutrons; therefore, they are more appropriate than thermoluminescent dosemeters in measuring neutron dose equivalent.<br /> (© The Author(s) 2020. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.)

Details

Language :
English
ISSN :
1742-3406
Volume :
188
Issue :
4
Database :
MEDLINE
Journal :
Radiation protection dosimetry
Publication Type :
Academic Journal
Accession number :
32025737
Full Text :
https://doi.org/10.1093/rpd/ncaa001