Does the magnetization transfer effect bias chemical exchange saturation transfer effects? Quantifying chemical exchange saturation transfer in the presence of magnetization transfer.

Authors :
Smith AK
Ray KJ
Larkin JR
Craig M
Smith SA
Chappell MA
Source :
Magnetic resonance in medicine [Magn Reson Med] 2020 Sep; Vol. 84 (3), pp. 1359-1375. Date of Electronic Publication: 2020 Feb 18.
Publication Year :
2020

Abstract

Purpose: Chemical exchange saturation transfer (CEST) is an MRI technique sensitive to the presence of low-concentration solute protons exchanging with water. However, magnetization transfer (MT) effects also arise when large semisolid molecules interact with water, which biases CEST parameter estimates if quantitative models do not account for macromolecular effects. This study establishes under what conditions this bias is significant and demonstrates how using an appropriate model provides more accurate quantitative CEST measurements.<br />Methods: CEST and MT data were acquired in phantoms containing bovine serum albumin and agarose. Several quantitative CEST and MT models were used with the phantom data to demonstrate how underfitting can influence estimates of the CEST effect. CEST and MT data were acquired in healthy volunteers, and a two-pool model was fit in vivo and in vitro, whereas removing increasing amounts of CEST data to show biases in the CEST analysis also corrupts MT parameter estimates.<br />Results: When all significant CEST/MT effects were included, the derived parameter estimates for each CEST/MT pool significantly correlated (P < .05) with bovine serum albumin/agarose concentration; minimal or negative correlations were found with underfitted data. Additionally, a bootstrap analysis demonstrated that significant biases occur in MT parameter estimates (P < .001) when unmodeled CEST data are included in the analysis.<br />Conclusions: These results indicate that current practices of simultaneously fitting both CEST and MT effects in model-based analyses can lead to significant bias in all parameter estimates unless a sufficiently detailed model is utilized. Therefore, care must be taken when quantifying CEST and MT effects in vivo by properly modeling data to minimize these biases.<br /> (© 2020 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine.)

Details

Language :
English
ISSN :
1522-2594
Volume :
84
Issue :
3
Database :
MEDLINE
Journal :
Magnetic resonance in medicine
Publication Type :
Academic Journal
Accession number :
32072677
Full Text :
https://doi.org/10.1002/mrm.28212