Back to Search Start Over

[Development of a Wearable Wireless ECG Monitoring System with Ultra-low Power Consumption].

Authors :
Sun Z
Ye J
Zhang X
Yuan M
Zhong Z
Tan X
Source :
Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation [Zhongguo Yi Liao Qi Xie Za Zhi] 2020 Jan 08; Vol. 44 (1), pp. 28-32.
Publication Year :
2020

Abstract

This study describes the development of a wireless and wearable ECG monitoring system with ultra-low power consumption. The system is mainly composed of a connection part of an ECG electrode sticker, an electrocardiogram collecting part, a data storage part, a Bluetooth main control unit, a charging module, a voltage regulator and a lithium battery. The low-power ECG acquisition chip ADS1292R and the ultra-low-power Bluetooth microcontroller nRF51822 together constitute the ECG signal acquisition and wireless data communication part. The collected ECG signals can be sent to the mobile APP through the Bluetooth function provided by the MCU, and can completly display and analysis to achieve low power system. After testing, the system power consumption is only (3.7 V×2.87 mA)10.619 mW, and if it is optimized, it can further reduce power consumption, therefore, the system design can have good applicability.

Details

Language :
Chinese
ISSN :
1671-7104
Volume :
44
Issue :
1
Database :
MEDLINE
Journal :
Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
Publication Type :
Academic Journal
Accession number :
32343062
Full Text :
https://doi.org/10.3969/j.issn.1671-7104.2020.01.006