Back to Search Start Over

Defibrillation threshold testing during implantable cardioverter defibrillator implantation: 5-year follow-up.

Authors :
Kim SS
Park HW
Jeong HK
Lee KH
Yoon NS
Cho JG
Source :
Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing [J Interv Card Electrophysiol] 2021 Apr; Vol. 60 (3), pp. 485-491. Date of Electronic Publication: 2020 May 13.
Publication Year :
2021

Abstract

Purpose: Defibrillation threshold (DFT) testing is a routine practice in some Asian countries for patients receiving an implantable cardioverter defibrillator (ICD). However, there are few long-term data about the necessity of intraoperative DFT testing in an Asian population. We investigated the safety of DFT testing and the long-term clinical outcomes in Asian patients undergoing ICD implantation.<br />Methods: All patients undergoing de novo transvenous ICD implantation were randomized to undergo periprocedural DFT testing. The study included 67 patients (50 males; 51.5 ± 16.9 years) who underwent ICD implantation with (n = 33) or without (n = 34) intraoperative DFT testing between March 2012 and February 2014. We compared first-shock success, composite safety end points (the sum of complications recorded at 30 days), arrhythmic death, and all-cause mortality.<br />Results: The baseline clinical characteristics and the procedural-related adverse event rate (3.0% with DFT vs. 0% with non-DFT, p = 0.214) did not differ between groups. The programmed output of the first shock was lower in the DFT testing group (22.9 ± 4.4 J vs. 25.3 ± 5.4 J, p = 0.007). However, there were no significant differences between groups for all-cause mortality (12.1% vs. 17.6%, p = 0.526) or first-shock success rate for ventricular arrhythmia (100% vs. 88.2%, p = 0.471).<br />Conclusions: There were no between-group differences in periprocedural safety, complications, and long-term clinical outcomes. Our results suggest that DFT testing in Asian patients allows reduction of the programmed output of the first shock, but does not affect long-term clinical outcomes.

Details

Language :
English
ISSN :
1572-8595
Volume :
60
Issue :
3
Database :
MEDLINE
Journal :
Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing
Publication Type :
Academic Journal
Accession number :
32399866
Full Text :
https://doi.org/10.1007/s10840-020-00733-x