Back to Search Start Over

Safety of leadless pacemaker implantation in the very elderly.

Authors :
Pagan E
Gabriels J
Khodak A
Chang D
Beldner S
Epstein LM
Willner J
Source :
Heart rhythm [Heart Rhythm] 2020 Dec; Vol. 17 (12), pp. 2023-2028. Date of Electronic Publication: 2020 May 23.
Publication Year :
2020

Abstract

Background: The Micra leadless pacemaker (MLP) has proven to be an effective alternative to a traditional transvenous pacemaker (TVP). However, there has been concern about using the MLP in frail elderly patients because of the size of the implant sheath and perceived risk of perforation.<br />Objectives: The objectives of this study were to report the safety of the MLP and compare MPLs with TVPs in the very elderly.<br />Methods: All patients 85 years and older who received an MLP or a single-chamber TVP across 6 hospitals in the Northwell Health system from December 2015 to November 2019 were included. Demographic characteristics, procedural details, and procedure-related complications were reviewed.<br />Results: Over 4 years, 564 patients underwent MLP implantation. During this time, 183 MLPs and 119 TVPs were implanted in patients 85 years and older. The mean age was 89.7 ± 3.4 years, and 47.4% were men. MLP implantation was successful in all but 3 patients (98.4% success rate). There was no difference in procedure-related complications (3.3% vs 5.9%; P = .276). Complications included 5 (2.7%) access site hematomas in the MLP group, 3 (2.5%) in the TVP group, 1 (0.5 vs 0.8%) pericardial effusion in each group, and 3 (2.5%) acute lead dislodgments (<24 hours) in the TVP group. MLP implantation resulted in a significantly shorter mean procedure time (35.7 ± 23.0 minutes vs 62.3 ± 31.5 minutes, P < .001).<br />Conclusion: In a large multicenter study of patients 85 years and older, MLP implantation (1) was successful in 98.4% of patients, (2) was safe with no difference in procedure-related complications compared to the TVP group, and (3) resulted in significantly shorter procedure times.<br /> (Published by Elsevier Inc.)

Details

Language :
English
ISSN :
1556-3871
Volume :
17
Issue :
12
Database :
MEDLINE
Journal :
Heart rhythm
Publication Type :
Academic Journal
Accession number :
32454218
Full Text :
https://doi.org/10.1016/j.hrthm.2020.05.022