Differences in clinical outcomes between pre- and post-marketing clinical study following paclitaxel-coated balloon catheter treatment for coronary in-stent restenosis: from the Japanese regulatory viewpoint.

Authors :
Mitsutake Y
Konishi A
Shiba T
Ito T
Ho M
Shirato H
Source :
Heart and vessels [Heart Vessels] 2021 Feb; Vol. 36 (2), pp. 155-162. Date of Electronic Publication: 2020 Aug 10.
Publication Year :
2021

Abstract

In 2013, a drug-coated balloon catheter (DCB) (SeQuent Please) for the treatment of coronary in-stent restenosis (ISR) was approved in Japan. The pre-marketing Japan domestic NP001 study demonstrated better outcomes of the DCB (n = 138) compared to plain balloon angioplasty (n = 72). After the introduction to marketing, a post-marketing surveillance (PMS) (n = 396) was conducted to evaluate the safety and efficacy of the DCB in Japanese routine clinical practice. The aim of this paper was to assess differences between the pre-marketing NP001 study and the PMS. Compared to the NP001 study, more complex lesions were treated in the PMS (type B2/C: 69.0% vs 20.4%, total occlusion: 11.2% vs 0%, p < 0.001, respectively) and target lesion was more frequently ISR related to drug-eluting stent (DES) (79.5% vs 39.4%, p < 0.001). Regarding clinical outcomes, the rate of target lesion revascularization (TLR) was higher in the PMS than in the NP001 study (TLR: 12.9% at 7 months and 17.6% at 12 months vs 2.8% at 6 months, p = 0.001, p < 0.001, respectively). Multivariable logistic regression analysis revealed that DES-ISR was a risk factor of TLR after DCB treatment for ISR (odds ratio: 5.77, 95% CI 1.75-18.95, p = 0.004). Among representative published trials using DCB for ISR, clinical outcomes are often worse in DES-ISR trials than those in bare metal stent-ISR trials. The rates of TLR in previous DES-ISR trials are similar to that in the current PMS (TLR at 12 months: 22.1% for ISAR-DESIRE 3, 15.3% for PEPCAD-DES, and 13.0% for RIBS IV). The effectiveness and safety of DCB for coronary ISR have been confirmed in the Japanese real-world survey. PMS would be useful to evaluate the safety and effectiveness of medical products throughout their total life cycles.

Details

Language :
English
ISSN :
1615-2573
Volume :
36
Issue :
2
Database :
MEDLINE
Journal :
Heart and vessels
Publication Type :
Academic Journal
Accession number :
32776235
Full Text :
https://doi.org/10.1007/s00380-020-01676-z