Imaging of two samples with a single transmit/receive channel using coupled ceramic resonators for MR microscopy at 17.2 T.

Authors :
Moussu MAC
Glybovski SB
Abdeddaim R
Craeye C
Enoch S
Tihon D
Kurdjumov S
Dubois M
Georget E
Webb AG
Belov P
Ciobanu L
Source :
NMR in biomedicine [NMR Biomed] 2020 Nov; Vol. 33 (11), pp. e4397. Date of Electronic Publication: 2020 Aug 31.
Publication Year :
2020

Abstract

In this paper we address the possibility to perform imaging of two samples within the same acquisition time using coupled ceramic resonators and one transmit/receive channel. We theoretically and experimentally compare the operation of our ceramic dual-resonator probe with a wire-wound solenoid probe, which is the standard probe used in ultrahigh-field magnetic resonance microscopy. We show that due to the low-loss ceramics used to fabricate the resonators, and a favorable distribution of the electric field within the conducting sample, a dual probe, which contains two samples, achieves an SNR enhancement by a factor close to the square root of 2 compared with a solenoid optimized for one sample.<br /> (© 2020 The Authors. NMR in Biomedicine published by John Wiley & Sons Ltd.)